Thyroid uptake and scintigraphy using 99mTc pertechnetate

standardization in normal individuals.

Autores

  • Celso Darío Ramos Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University
  • Denise Engelbrecht Zantut Wittmann Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University
  • Elba Cristina Sá de Camargo Etchebehere Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University
  • Marcos Antonio Tambascia Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University
  • Cleide Aparecida Moreira Silva Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University
  • Edwaldo Eduardo Camargo Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University

Palavras-chave:

Tireóide, Cintilografia, Captação, Pertecnetato- 99mTc, Valores normais

Resumo

CONTEXTO: A utilização do pertecnetato-99mTc para captação e cintilografia da tireóide exibe vantagens em relação ao iodeto-131I, como menor exposição do paciente à radiação, imagens de melhor qualidade e maior rapidez para a realização do procedimento. OBJETIVO: Padronizar uma metodologia simples e rápida para a realização de captação e cintilografia da tireóide e determinar os valores normais de captação tireoidiana de pertecnetato-99mTc. TIPO DE ESTUDO: Prospectivo, não-randomizado. LOCAL: Serviço de Medicina Nuclear, Faculdade de Ciências Médicas da Universidade Estadual de Campinas, Campinas, Brasil. PARTICIPANTES: Estudamos 47 indivíduos normais, 30 mulheres e 17 homens, com idades variando entre 19 e 61 anos (média 33 anos). PROCEDIMENTOS: A avaliação laboratorial da função tireoidiana foi feita por meio de dosagens séricas de tiroxina e tireotrofina. As imagens foram adquiridas 20 minutos após a injeção venosa de 10 mCi (370 MBq) de pertecnetato-99mTc em uma câmara de cintilação computadorizada equipada com colimador de baixa energia, alta resolução e furos paralelos. RESULTADOS: Todos os indivíduos eram clínica e laboratorialmente eutireoidianos. A captação tireoidiana basal de pertecnetato-99mTc variou de 0,4 a 1,7%. A análise estatística evidenciou que 95% dos indivíduos normais estudados apresentaram valores de captação entre 0,4 a 1,5% da dose injetada. CONCLUSÃO: A avaliação da estrutura e função tireoidianas com o uso do pertecnetato-99mTc é um método simples, rápido e eficiente, podendo ser facilmente incorporado à rotina dos serviços de medicina nuclear.

Downloads

Não há dados estatísticos.

Biografia do Autor

Celso Darío Ramos, Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University

MD. Division of Nuclear Medicine, Department of Radiology, School of Medical Sciences, Campinas State University, Campinas, Brazil.

Denise Engelbrecht Zantut Wittmann, Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University

MD, PhD. Division of Endocrinology, Department of Internal Medicine, School of Medical Sciences, Campinas State University, Campinas, Brazil.

Elba Cristina Sá de Camargo Etchebehere, Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University

MD. Division of Nuclear Medicine, Department of Radiology, School of Medical Sciences, Campinas State University, Campinas, Brazil.

Marcos Antonio Tambascia, Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University

MD, PhD. Division of Endocrinology, Department of Internal Medicine, School of Medical Sciences, Campinas State University, Campinas, Brazil.

Cleide Aparecida Moreira Silva, Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University

Statistician, Research Commission, School of Medical Sciences, Campinas State University, Campinas, Brazil.

Edwaldo Eduardo Camargo, Division of Nuclear Medicine; Division of Endocrinology; Statistical Nucleus; School of Medical Sciences, Campinas State University

MD, PhD. Full Professor, Division of Nuclear Medicine, Department of Radiology, School of Medical Sciences, Campinas State University, Campinas, Brazil.

Referências

Chapman EM. History of the discovery and early use of radioactive iodine. JAMA 1983;15:2042-4.

Cavalieri RR, McDougall IR. “In vivo” isotopic tests and imaging. In: Braverman LE, Utiger R. Werner & Ingbar. The thyroid. 7th ed. Philadelphia: Lippincott-Ranen; 1996:1-372.

Becker D, Charkes ND, Dworkin H, et al. Procedure guideline for thyroid scintigraphy. J Nucl Med 1996;37:1264-6.

Paras P, Hamilton DR, Evans C, Herrera NE, Lagunas Solar MC. Iodine-123 assay using a radionuclide calibrator. Int J Nucl Med Biol 1983;10:111-5.

Ziessman HA, Fahey FH, Gochoco JM. Impact of radiocontaminants in commercially available iodine-123: dosimetric evaluation. J Nucl Med 1986;27:428-32.

Burke G, Halko A, Silverstein GE, Hilligoss M. Comparative thyroid uptake studies with 131I and 99mTcO . J Clin Endocrinol Metab 1972;34:630-7.

Ramos CD, Zantut Wittmann DE, Tambascia MA, Assumpção L, Etchebehere EC, Camargo EE. Thyroid suppression test with L-thyroxine and [99mTc] pertechnetate. Clin Endocrinol 2000;52:471-7.

Higgins HP, Ball D, Eastham S. 20-Min 99mTc thyroid uptake: a simplified method using the gamma camera. J Nucl Med 1973;14:907-11.

Andros G, Harper PV, Lathrop KA, McCardle RJ. Pertechnetate- 99m localization in man with applications to thyroid scanning and the study of thyroid physiology. J Clin Endocrinol Metab 1965; 25:1067-76.

Smith JJ, Croft BY, Brookeman VA, Teates CD. Estimation of 24- hour thyroid uptake of I-131 sodium iodide usinga 5-minute uptake of technetium-99m pertechnetate. Clin Nucl Med 1990;15:80-3.

Sodee DB. The study of thyroid physiology utilizing intravenous sodium pertechnetate. J Nucl Med 1966;7:564-7.

Maisey MN, Natarajan TK, Hurley PJ, Wagner HN Jr. Validation of a rapid computerized method of measuring99mTc pertechnetate uptake for routine assessment of thyroid structure and function. J Clin Endocrinol Metab 1973;36:317-22.

Schneider PB. Simple, rapid thyroid function testing with 99mTc- pertechnetate thyroid uptake ratio and neck/thigh ratio. Am J Roentgenol 1979;132:249-353.

Selby JB, Buse MG, Gooneratne NS, Moore DO. The Anger camera and the pertechnetate ion in the routine evaluation of thyroid uptake and imaging. Clin Nucl Med 1979;4:233-7.

Harbert JC. The thyroid. In: Harbert JC, Eckelman WC, Neumann RD. Nuclear medicine: diagnosis and therapy. New York: Thieme Medical Publishers; 1996:407-27.

Hurley PJ, Maisey MN, Natarajan TK, Wagner HN Jr. A computerized system for rapid evaluation of thyroid function. J Clin Endocrinol Metab 1972;34:354-60.

Sucupira MS, Camargo EE, Nickoloff EL, Alderson PO, Wagner HN. The role of 99mTc pertechnetate uptake in the evaluation of thyroid function. Int J Nucl Med Biol 1983;10:29-33.

Downloads

Publicado

2002-03-03

Como Citar

1.
Ramos CD, Wittmann DEZ, Etchebehere ECS de C, Tambascia MA, Silva CAM, Camargo EE. Thyroid uptake and scintigraphy using 99mTc pertechnetate: standardization in normal individuals. Sao Paulo Med J [Internet]. 3º de março de 2002 [citado 16º de outubro de 2025];120(2):45-8. Disponível em: https://periodicosapm.emnuvens.com.br/spmj/article/view/2685

Edição

Seção

Artigo Original